Infrared Spectrophotometer, Spectrophotometers (RAY-30)
The RAY-30 Infrared Oil Meter can be used to record either IR absorption or (with accessories) IR reflection spectra in the range of 4000 cm-1 to 400 cm-1 (wavelengths from 2.5 um to 25um). The user can do
spectrum. It is a useful tool for petrology, the chemical industry, medical work, food,safety, the paper industry, environmental monitoring, the semiconductor industry, etc.,for material structure analysis.
2 Main performance index1.1 Wavenumber range: 4000 cm-1 to 400 cm-1
1.2 Transmittance scale: 0 to 100% (expandable to -400% to 400%)
1.3 Absorbance scale: 0 to 1A (expandable to -4A to 4A)
1.4 Wavenumber scale expansion: 4000cm-1~400cm-1 (optional)
1.5 Transmittance scale expansion: -400%~400%(optional)
    Single beam scale expansion: -400%~400%(optional)
1.6 Scan time for entire range: approx. 2.5 min. to 25 min (in fast response mode)
1.7 Slit program: 5 slit widths: Very Wide, Wide, Normal, Narrow, Very Narrow
1.8 Dimensions: 800mm×610mm×300mm
1.9 Weight: approx. 100 kg
2.1 Wavenumber accuracy: +-4cm-1(4000-2000cm-1);+-2cm-1(2000-400cm-1)
2.2 Wavenumber reproducibility: 2cm-1(4000-2000cm-1);1cm-1(2000-400cm-1)
2.3 Transmittance accuracy: +-0.2% (not including noise)
2.4Transmittance reproducibility: less than 0.4% (1000-930cm-1)
2.5Baseline flatness before correction, less than +-4%,
2.6 Resolution: for polystyrene, having 6 absorption peaks near 3000 cm-1;sixth peak has height above 2%; for ammonia gas near 1000 cm-1, resolution
is 1.5cm-1; peak height is above 0.5%
2.7Scattered light coefficient: less than 0.5% (4000-650cm-1); less than 1%(650-400cm-1)
2.8 Main control and data processing functions: 0 to 100 %T automatic calibration, horizontal and vertical scale expansion, Io baseline shift,wavenumber reset, wavenumber check, automatic filter change, slit
programming, automatic gain adjust, repetitive scan, print spectral peak values, spectral data printout, spectrum derivative (first derivative, second derivative), spectral data smoothing, absorption spectrum expansion,spectrum accumulate, mathematical operations between spectra (+, -, x, /),baseline correction, transmittance to absorbance conversion, spectral file management, (read, save, delete, etc)